Features:- Simultaneously measures and records pulse signals, events, state changes, and runtimes.- Stores up to over 4 million measurements, enabling longer deployments with fewer site visits.- Maximum pulse rate: 120 Hz.- Pulse input range: 0 to 24VDC.-
Features:
– Simultaneously measures and records pulse signals, events, state changes, and runtimes.
– Stores up to over 4 million measurements, enabling longer deployments with fewer site visits.
– Maximum pulse rate: 120 Hz.
– Pulse input range: 0 to 24VDC.
– Streamlines deployment via a range of start/stop options, logger status LEDs, and high-speed USB 2.0 data offload.
– Works with Onset’s Energy & Power Meter to measure Power Factor (PF), Reactive Power (VAR), Watt Hours (Wh) and more. Learn how.
– The T-VER-E50B2 Energy and Power Meter outputs (3) sets of pulses which are logged by the UX120-017. These pulses represent Watt-hours, Amp-hours, andVAR-hours. HOBOware software uses these pulse values to calulate AC Current, AC Voltage, kW, Power Factor, VARs, and VA. The formulas used in these calculationscan be found here: 14993-A-Derived-Channels.zip.
– Compatible with HOBOware and HOBOware Pro software for logger setup, graphing and analysis.
– External Contact Input: Electronic solid state switch closure or logic driven digital signals to 24 V
– Maximum Pulse Frequency: 120 Hz
– Maximum State, Event, Runtime Frequency: 1 Hz
– Bits: 4 to 32 bits depending on pulse rate and logging interval
– Maximum Pulses Per Interval: 7,863,960 (using maximum logging rate)
– Driven Logic Signal: Input Low: less than or equal to 0.4 V; Input High: 3 to 24 V
– Absolute Maximum Rating: Maximum Voltage: 25 V DC
– v Minimum Voltage: -0.3 V DC
– Solid State Switch Closure: Input Low: less than 10 K?; Input High: more than 500 K?
– Internal Weak Pull-Up: 100 K?
– Input Impedance: Solid state switch closure: 100 K? pull up; Driven signal: 4.5 K?
– Minimum Pulse Width: Contact closure duration: 500 uS; Driven logic signal: 100 uS
– Lockout Time: 0 to 1 second in 100 ms steps
– Edge Detection:Falling edge, Schmitt Trigger buffer
– Preferred Switch State: Normally open or Logic “1” state
– Resolution: Pulse: 1 pulse, Runtime: 1 second, State and Event: 1 State or Event
– Logging Rate:1 second to 18 hours, 12 minutes, 15 seconds
– Memory: 520,192 measurements (assumes 8-bit)
– Download Type: USB 2.0 interface
– Download Time: 30 seconds for UX120-017, 1.5 minutes for UX120-017M
– Time Accuracy: +- 1 minute per month at 25 C (77 F) (see Plot A on next page)
– Battery Type: Two AA alkaline or lithium batteries
– Battery Life:1 year, typical with logging intervals greater than 1 minute and normallyopen contacts
Supported measurements:
The UX120-017 data logger supports the following measurements: AC Voltage, Amp Hour (Ah), Amps (A), Event, Kilowatt Hours (kWh), Kilowatts (kW), MotorOn/Off, Power Factor (PF), Pulse Input, Runtime, State Open/Closed, Volt-Amp Reactive, Volt-Amp Reactive hour, Volt-Amps (VA), Volts (V), Watt Hours (Wh)and Watts (W).
$25
We offer free shipping on orders over $30. Please check the free - shipping eligibility at checkout.
It usually takes [3-5] business days for standard shipping. Please note that this is an estimated time frame and may be affected by local holidays, and unforeseen circumstances.